|
Main /
Calibration menuThe calibration menu can be used to update or reset the level calibration of the tinySA
Calibration is used to calibrate the power measurement levels. Before starting the calibration, connect the low and high input/output using a short SMA cable. After connecting the cable, activate CALIBRATE. The calibration can be reset by RESET CALIBRATION. At the end of the calibration switch off the tinySA. After switching on again the numbers indicating the power levels should be in white to indicate the power reading is calibrated. An uncalibrated state is indicated with red level indicators. The level calibration data is stored in the leveloffset configuration variable. |