Allows entering the actual levels of either input or output
For input level calibration you need a calibrated signal generator for the range you are interested in.
For output level calibration you need a calibrated power meter for the range you are interested in.
- INPUT LEVEL allows calibration of the dBm reading by entering the know level of the signal under marker 1. The updated calibration value is stored in the leveloffset configuration variable. Power calibration is removed by setting the value to 100. Setting the actual power sets the leveloffset for all frequencies. A detailed power level correction versus frequency is available using the correction command
- OUTPUT LEVEL allows entering of the actual level of the automatically activated 30MHz output
- BACK returns to the Expert Config menu