The configuration menu can be used to update various settings and to test or calibrate the tinySA
- TOUCH activates the touch sub menu
- TOUCH CAL enables the calibration of the touch panel. The results are stored in NVM.
- TOUCH TEST is used to verify touch calibration. (Touch and move the stylus to see that it draws on screen.)
- SELF TEST is used after a cable is used to connect the two SMA connectors. The progress is displayed and you need to touch the display after each failing test and at the end of the test.
- LEVEL CAL activates the calibration menu.
- VERSION shows the SW version information.
- SPUR REMOVAL? activates the two spur removal mechanisms. Below 190MHz the above and below IF are used to identify spurs and above 190MHZ the IF is shifted.
- SAMPLE REP sets the amount of samples to take at a frequency. Minimum is 1, maximum is not defined but a large number will make the scan very slow. Increasing the sample repeat helps to average out noise.)
- BRIGHTNESS allows setting the brightness of the screen using the leveler button.
- DATE TIME activates the menu for setting Date and Time. (The battery provides power to store the Date and Time, once set.)
- -> TinySA4/MORE? activates the next configuration menu
- BACK moves back to the input menu.